Product Description
Documentation
- Simulate output to discover malfunctions
- Simulate temperature
- Provided with universal BNC connector
Sometimes it is difficult to recognize whether a particular malfunction is due to the meter or the electrode. By simply connecting HI 931001 to your meter's input socket and turning the dials, pH readings can be simulated from 0 to 14 pH in 0.01 steps. The output signals all correspond to pH values at 25°C and your pH meter should be able to check the span of your pH meter.
For the mV range, HI 931001 can simulate output from -1000 to +1000 mV in 1 mV steps.
Specifications:
Range: pH 0.00 to 14.00 pH
ORP -1000 to 1000
Resolution: pH 0.01 pH
ORP 1 mV
Accuracy @ 20°C: pH ±0.01 pH
ORP ±1 mV
Temperature:
Compensation: all output values are simulated at 25°C
Battery Type / Life: 9V/approximately 500 hours of use
Environment: 0 to 50°C (32 to 122°F); RH max 95%
Dimensions: 185 x 82 x 53 mm (7.3 x 3.2 x 2.1’’)
Weight: 320 g (11.3 oz.)
Order Information:
HI 931001 is supplied with HI 7858/1 BNC/BNC coaxial cable